Abstract

Assessment of population exposure to VOC in ambient atmospheres is receiving heightened interest as the adverse health effects of chronic exposure to certain of these compounds are identified. Active (pumped) and passive samplers are the most commonly used devices for this type of monitoring. It has been shown, however, that these devices, along with all other preconcentration techniques, are susceptible to ozone interference. It is demonstrated that this interference occurs even at low ozone concentrations and that it may result in the under-estimation of population exposure. A convenient and effective ozone scrubbing method is identified and successfully applied and validated for both active and passive samplers for a range of VOC

Dates and versions

ineris-00962773 , version 1 (21-03-2014)

Identifiers

Cite

Matthew S. Bates, Norbert Gonzalez-Flesca, Ranjeet Sokhi, Vincenzo Cocheo. Atmospheric volatile organic compounds monitoring Ozone induced artefact formation. Environmental Monitoring and Assessment, 2000, 65 (1-2), pp.89-97. ⟨10.1023/A:1006420412523⟩. ⟨ineris-00962773⟩