Evaluation de l'efficacité de prélèvement d'un précipitateur électrostatique

Abstract : Easy-to-use particle sampling techniques leading to easy-to-analyze samples is a growing need for nanosafety. Electrostatic precipitation [Dickens 1999] commercialized by TSI (Nanometer Aerosol Sampler model 3089) could be an answer. However, more information is needed on the performance of this instruments. On the one hand, real time sampling efficiency of the NAS regarding voltage has been studied. On the other hand, the sampling efficiency for 80 nm polystyrene particles has been considered through both real time CPC measurements and TEM analysis: these methods lead to very different results, respectively 62-70% and < 0,1 %. These results are discussed regarding literature
Document type :
Conference papers
Complete list of metadatas

Cited literature [1 references]  Display  Hide  Download

https://hal-ineris.archives-ouvertes.fr/ineris-00973609
Contributor : Gestionnaire Civs <>
Submitted on : Friday, April 4, 2014 - 2:11:55 PM
Last modification on : Friday, April 4, 2014 - 2:11:55 PM
Long-term archiving on : Friday, July 4, 2014 - 12:26:55 PM

File

2011-027_hal.pdf
Files produced by the author(s)

Identifiers

  • HAL Id : ineris-00973609, version 1
  • INERIS : EN-2011-027

Collections

Citation

Olivier Le Bihan, Aurélien Zantman, S. Thevenet, Badr R'Mili, Aurélien Ustache, et al.. Evaluation de l'efficacité de prélèvement d'un précipitateur électrostatique. 26. Congrès Français sur les Aérosols (CFA 2011), Jan 2011, Paris, France. pp.NC. ⟨ineris-00973609⟩